High-frequency optical coherence tomography predictors of aneurysm occlusion following flow diverter treatment in a preclinical model

King, Robert M
Peker, Ahmet
Anagnostakou, Vania
Raskett, Christopher M
Arends, Jennifer M
Dixit, Harish G
Ughi, Giovanni J
Puri, Ajit S
Gounis, Matthew J
Shazeeb, Mohammed Salman
Embargo Expiration Date
Abstract

Background: High-frequency optical coherence tomography (HF-OCT) is an intravascular imaging method that allows for volumetric imaging of flow diverters in vivo.

Objective: To examine the hypothesis that a threshold for both volume and area of communicating malapposition can be predictive of early aneurysm occlusion.

Methods: Fifty-two rabbits underwent elastase aneurysm formation, followed by treatment with a flow diverter. At the time of implant, HF-OCT was acquired to study the rate and degree of communicating malapposition. Treated aneurysms were allowed to heal for either 90 or 180 days and euthanized following catheter angiography. Healing was dichotomized into aneurysm remnant or neck remnant/complete occlusion. Communicating malapposition was measured by HF-OCT using a semi-automatic algorithm able to detect any points where the flow diverter was more than 50 µm from the vessel wall. This was then summed across image slices to either a volume or area. Finally, a subsampled population was used to train a statistical classifier for the larger dataset.

Results: No difference in occlusion rate was found between device type or follow-up time (p=0.28 and p=0.67, respectively). Both volume and area of malapposition were significantly lower in aneurysms with a good outcome (p<0.001, both). From the statistical model, a volume of less than 0.56 mm3 or a normalized area less than 0.69 as quantified by HF-OCT was predictive of occlusion (p<0.001, each).

Conclusions: HF-OCT allows for measurements of both volume and area of malapposition and, from these measurements, an accurate prediction for early aneurysm occlusion can be made.

Source

King RM, Peker A, Anagnostakou V, Raskett CM, Arends JM, Dixit HG, Ughi GJ, Puri AS, Gounis MJ, Shazeeb MS. High-frequency optical coherence tomography predictors of aneurysm occlusion following flow diverter treatment in a preclinical model. J Neurointerv Surg. 2022 Aug 24:jnis-2022-019275. doi: 10.1136/jnis-2022-019275. Epub ahead of print. PMID: 36002288.

Year of Medical School at Time of Visit
Sponsors
Dates of Travel
DOI
10.1136/jnis-2022-019275
PubMed ID
36002288
Other Identifiers
Notes
Funding and Acknowledgements
Corresponding Author
Related Resources
Related Resources
Repository Citation
Rights
© Author(s) (or their employer(s)) 2022. No commercial re-use. See rights and permissions. Published by BMJ.
Distribution License