A complication involving a braided hook-wire localization device
D'Orsi, Carl J. ; Swanson, Richard S. ; Moss, Lawrence J. ; Reale, Frank R. ; Wertheimer, Michael D.
D'Orsi, Carl J.
Swanson, Richard S.
Moss, Lawrence J.
Reale, Frank R.
Wertheimer, Michael D.
Citations
Altmetric:
Student Authors
Faculty Advisor
Academic Program
UMass Chan Affiliations
Document Type
Journal Article
Publication Date
1993-05-01
Subject Area
Embargo Expiration Date
Link to Full Text
Abstract
At the authors' institution, needle localization of breast lesions with a braided hook wire involves the wire being cut 1-2 cm from the point of entry before dissection, to avoid contamination of the sterile field with the nonsterile portion of wire. During dissection, the wire is brought through the skin into the area of dissection. In one patient, fragments of wire filaments were left within the breast. Braided hook wires must be cut cleanly, the cut surface should be wiped before dissection, and the surgical area should be cleansed before closure.