BACKGROUND: Digital chest drainage systems (DCDS) provide reliable pleural drainage while quantifying fluid output and air leak. However, the benefits of DCDS in the contemporary era of minimally invasive thoracic surgery and enhanced recovery after surgery (ERAS) protocols have not been fully investigated. Additionally, hospital and resident staff experiences after implementation of a DCDS have not been fully explored. The objective of this study was to evaluate the clinical outcomes and hospital staff experience after adoption of a DCDS for minimally invasive lung resections.

METHODS: A single-center retrospective review of patients who underwent minimally invasive lung resection (lobectomy, segmentectomy, and wedge resection) and received a DCDS from 11/1/2021 to 11/1/2022. DCDS patients were compared to sequential historical controls (3/1/2019-6/30/2021) who received a analog chest drainage system. For the analog system, chest tubes were removed when no bubbles were observed in the water seal compartment with Valsalva, cough, and in variable positions. With a DCDS, chest tubes were removed when the air leak was less than 30 cc/min for 8 hours, with no spikes. All patients followed an institutional ERAS protocol. Primary outcomes were length of stay (LOS) and chest tube duration. Hospital staff and residents were surveyed regarding their experience.

RESULTS: One hundred and twenty-four patients received DCDS, and 248 received an analog chest drainage system. There was a reduction in mean LOS (3.6 . 4.4 days, P=0.01) and chest tube duration (2.7 . 3.6 days, P=0.03) in the DCDS group. Hospital staff (n=77, 46% response rate) reported the DCDS easier to use (60%, P<0.001) and easier to care for patients with (65%, P<0.001) compared to the analog system. Surgical residents (n=28, 56% response rate) reported increased confidence in interpretation of air leak (75%, P<0.001) and decision-making surrounding chest tube removal (79%, P<0.001).

CONCLUSIONS: Using a DCDS can reduce LOS and chest tube duration in the contemporary setting of minimally invasive lung resections and ERAS protocols. Increased confidence of resident decision-making for chest tube removal may contribute to improved outcomes.

Source

Palleiko BA, Singh A, Strader C, Patil T, Crawford A, Emmerick I, Lou F, Uy K, Maxfield MW. Clinical outcomes and staff satisfaction after adoption of digital chest drainage system for minimally invasive lung resections. J Thorac Dis. 2024 May 31;16(5):2963-2974. doi: 10.21037/jtd-23-1747. Epub 2024 May 16. PMID: 38883637; PMCID: PMC11170378.

Year of Medical School at Time of Visit
Sponsors
Dates of Travel
DOI
10.21037/jtd-23-1747
PubMed ID
38883637
Other Identifiers
Notes
Funding and Acknowledgements
Corresponding Author
Related Resources
Related Resources
Repository Citation
Rights
Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non- commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0/