Loading...
Catastrophic tibial baseplate failure of a modern cementless total knee arthroplasty implant
Scully, William F. ; Deren, Matthew E. ; Bloomfield, Michael R.
Scully, William F.
Deren, Matthew E.
Bloomfield, Michael R.
Citations
Altmetric:
Student Authors
Faculty Advisor
Academic Program
UMass Chan Affiliations
Document Type
Journal Article
Publication Date
2019-10-16
Subject Area
Embargo Expiration Date
Link to Full Text
Abstract
Tibial baseplate fracture following primary total knee arthroplasty is a rare complication, particularly with modern implants and surgical techniques. This case details the first known report of mid-range follow-up catastrophic failure of a cementless modular, trabecular metal tibial baseplate. This failure highlights the importance of continued follow-up for novel implants, to include cementless knee arthroplasty designs, particularly if new symptoms arise or periarticular bone loss is identified on radiograph. and Knee Surgeons.
Source
Arthroplast Today. 2019 Oct 16;5(4):446-452. doi: 10.1016/j.artd.2019.09.002. eCollection 2019 Dec. Link to article on publisher's site
Year of Medical School at Time of Visit
Sponsors
Dates of Travel
DOI
10.1016/j.artd.2019.09.002
Permanent Link to this Item
PubMed ID
31886388
Other Identifiers
Notes
Funding and Acknowledgements
Corresponding Author
Related Resources
Related Resources
Repository Citation
Rights
© 2019 Published by Elsevier Inc. on behalf of The American Association of Hip and Knee Surgeons. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).