Background: Treatment of large vessel occlusion (LVO) using mechanical thrombectomy with or without intravenous thrombolysis has demonstrated better outcomes compared to medical treatment alone. Large-bore aspiration catheters have been recently introduced. Their effectiveness and safety have not been demonstrated in a randomized trial. The SUMMIT MAX study is designed to address this question.
Methods: SUMMIT MAX is a randomized controlled trial where the effectiveness and safety of the large-bore Monopoint Reperfusion system (Route 92 Medical, San Mateo, CA), will be compared to the currently largest available FDA-cleared aspiration thrombectomy device the AXS Vecta Aspiration system (Stryker Neurovascular, Fremont, CA). The study is a multi-center, prospective, randomized, controlled, interventional, open label clinical trial. The hypothesis is that the effectiveness measured by the recanalization rate (modified thrombolysis in cerebrovascular infarction - mTICI) and safety measured by symptomatic intracranial hemorrhage rate (sICH) of the medical monopoint reperfusion system is non-inferior to the AXS Vecta Aspiration system.
Results: Up to 250 subjects are enrolled with at least 50% of subjects enrolled by US sites. The primary effectiveness endpoint is successful arterial revascularization defined as an mTICI score ≥ 2b after use of the assigned device adjudicated by an independent core lab. The primary safety endpoint is defined as sICH within 24 h (-8/+24) post-procedure. Secondary endpoints include successful arterial revascularization defined as a mTICI score ≥ 2b after use of the assigned device with or without adjunctive therapy; device-related serious adverse events; all asymptomatic hemorrhages; time from groin puncture to final angiogram; and rate of first pass effect defined as mTICI 2b after first pass with the assigned device stratified by age (≤85, ≥ 86).
Conclusion: SUMMIT MAX is a randomized controlled trial comparing the effectiveness and safety of a new large bore class of aspiration devices to the currently largest FDA-cleared aspiration device available.
Source
Dabus G, Puri AS, McGuinness B, Priest RA, Rai AT, Gross BA, Zaidat OO, Hanel RA, Shazam Hussain M, Shaikh HA, English JD, Nguyen TN. A prospective, randomized, controlled, interventional clinical trial to evaluate the safety and efficacy of the medical monopoint reperfusion system for aspiration thrombectomy in acute ischemic stroke patients (SUMMIT MAX): Trial rationale and design. Interv Neuroradiol. 2024 Sep 9:15910199241270711. doi: 10.1177/15910199241270711. Epub ahead of print. PMID: 39246132.