High-throughput multiplexed T-cell-receptor excision circle quantitative PCR assay with internal controls for detection of severe combined immunodeficiency in population-based newborn screening
Gerstel-Thompson, Jacalyn L. ; Wilkey, Jonathan F. ; Baptiste, Jennifer C. ; Navas, Jennifer S. ; Pai, Sung-Yun ; Pass, Kenneth A. ; Eaton, Roger B. ; Comeau, Anne Marie
Citations
Student Authors
Faculty Advisor
Academic Program
UMass Chan Affiliations
Document Type
Publication Date
Keywords
Calibration
DNA
Feasibility Studies
Gene Dosage
Genes, T-Cell Receptor
Humans
Infant, Newborn
Intensive Care Units
*Neonatal Screening
Polymerase Chain Reaction
Quality Control
Receptors, Antigen, T-Cell
Regression Analysis
Ribonucleases
Severe Combined Immunodeficiency
Genetics and Genomics
Medical Genetics
Pediatrics
Subject Area
Embargo Expiration Date
Link to Full Text
Abstract
BACKGROUND: Real-time quantitative PCR (qPCR) targeting a specific marker of functional T cells, the T-cell-receptor excision circle (TREC), detects the absence of functional T cells and has a demonstrated clinical validity for detecting severe combined immunodeficiency (SCID) in infants. There is need for a qPCR TREC assay with an internal control to monitor DNA quality and the relative cellular content of the particular dried blood spot punch sampled in each reaction. The utility of the qPCR TREC assay would also be far improved if more tests could be performed on the same newborn screening sample.
METHODS: We approached the multiplexing of qPCR for TREC by attenuating the reaction for the reference gene, with focus on maintaining tight quality assurance for reproducible slopes and for prevention of sample-to-sample cross contamination. Statewide newborn screening for SCID using the multiplexed assay was implemented, and quality-assurance data were recorded.
RESULTS: The multiplex qPCR TREC assay showed nearly 100% amplification efficiency for each of the TREC and reference sequences, clinical validity for multiple forms of SCID, and an analytic limit of detection consistent with prevention of contamination. The eluate and residual ghost from a 3.2-mm dried blood spot could be used as source material for multiplexed immunoassays and multiplexed DNA tests (Multiplex Plus), with no disruption to the multiplex TREC qPCR.
CONCLUSIONS: Population-based SCID newborn screening programs should consider multiplexing for quality assurance purposes. Potential benefits of using Multiplex Plus include the ability to perform multianalyte profiling.
Source
Clin Chem. 2010 Sep;56(9):1466-74. Epub 2010 Jul 21. Link to article on publisher's site