Calibration and optimization of 3D digital breast tomosynthesis guided near infrared spectral tomography
Michaelsen, Kelly E. ; Krishnaswamy, Venkataramanan ; Shi, Linxi ; Vedantham, Srinivasan ; Poplack, Steven P. ; Karellas, Andrew ; Pogue, Brian W. ; Paulsen, Keith D.
Citations
Student Authors
Faculty Advisor
Academic Program
UMass Chan Affiliations
Document Type
Publication Date
Subject Area
Collections
Embargo Expiration Date
Link to Full Text
Abstract
Calibration of a three-dimensional multimodal digital breast tomosynthesis (DBT) x-ray and non-fiber based near infrared spectral tomography (NIRST) system is challenging but essential for clinical studies. Phantom imaging results yielded linear contrast recovery of total hemoglobin (HbT) concentration for cylindrical inclusions of 15 mm, 10 mm and 7 mm with a 3.5% decrease in the HbT estimate for each 1 cm increase in inclusion depth. A clinical exam of a patient's breast containing both benign and malignant lesions was successfully imaged, with greater HbT was found in the malignancy relative to the benign abnormality and fibroglandular regions (11 muM vs. 9.5 muM). Tools developed improved imaging system characterization and optimization of signal quality, which will ultimately improve patient selection and subsequent clinical trial results.
Source
Biomed Opt Express. 2015 Nov 19;6(12):4981-91. doi: 10.1364/BOE.6.004981. eCollection 2015. Link to article on publisher's site