Publication

Dosimetric properties of high energy current (HEC) detector in keV x-ray beams

Zygmanski, Piotr
Shrestha, Suman
Elshahat, Bassem
Karellas, Andrew
Sajo, Erno
Citations
Altmetric:
Student Authors
Faculty Advisor
Academic Program
UMass Chan Affiliations
Document Type
Journal Article
Publication Date
2015-04-07
Subject Area
Embargo Expiration Date
Abstract

We introduce a new x-ray radiation detector. The detector employs high-energy current (HEC) formed by secondary electrons consisting predominantly of photoelectrons and Auger electrons, to directly convert x-ray energy to detector signal without externally applied power and without amplification. The HEC detector is a multilayer structure composed of thin conducting layers separated by dielectric layers with an overall thickness of less than a millimeter. It can be cut to any size and shape, formed into curvilinear surfaces, and thus can be designed for a variety of QA applications. We present basic dosimetric properties of the detector as function of x-ray energy, depth in the medium, area and aspect ratio of the detector, as well as other parameters. The prototype detectors show similar dosimetric properties to those of a thimble ionization chamber, which operates at high voltage. The initial results obtained for kilovoltage x-rays merit further research and development towards specific medical applications.

Source

Phys Med Biol. 2015 Apr 7;60(7):N121-9. 2015 Mar 19. Link to article on publisher's site

Year of Medical School at Time of Visit
Sponsors
Dates of Travel
DOI
10.1088/0031-9155/60/7/N121
PubMed ID
25789488
Other Identifiers
Notes
Funding and Acknowledgements
Corresponding Author
Related Resources
Related Resources
Repository Citation
Rights
Distribution License