Safety and effectiveness assessment of the surpass evolve (SEASE): a post-market international multicenter study
Vivanco-Suarez, Juan ; Dibas, Mahmoud ; Lopes, Demetrius Klee ; Hanel, Ricardo A ; Martínez-Galdámez, Mario ; Rodriguez-Calienes, Aaron ; Cortez, Gustavo M ; Fifi, Johanna T ; Devarajan, Alex ; Toth, Gabor ... show 10 more
Citations
Authors
Dibas, Mahmoud
Lopes, Demetrius Klee
Hanel, Ricardo A
Martínez-Galdámez, Mario
Rodriguez-Calienes, Aaron
Cortez, Gustavo M
Fifi, Johanna T
Devarajan, Alex
Toth, Gabor
Patterson, Thomas
Altschul, David
Pereira, Vitor M
Liu, Eileen
Puri, Ajit S
Kuhn, Anna Luisa
Guerrero, Waldo R
Khandelwal, Priyank
Bach, Ivo
Kan, Peter
Edhayan, Gautam
Given, Curtis
Narayanan, Sandra
Gross, Bradley A
Farooqui, Mudassir
Galecio-Castillo, Milagros
Derakhshani, Shahram
Ortega-Gutierrez, Santiago
Student Authors
Faculty Advisor
Academic Program
UMass Chan Affiliations
Document Type
Publication Date
Keywords
Subject Area
Collections
Embargo Expiration Date
Link to Full Text
Abstract
Background: Flow diverters are the first-line treatment for specific intracranial aneurysms (iA). Surpass Evolve (SE) is a new-generation 64-wire flow diverter with a high braid angle. Current literature on the SE is limited. We aimed to report the first international real-world experience evaluating the safety and effectiveness of the SE.
Methods: The Safety and Effectiveness Assessment of the Surpass Evolve (SEASE) was a multicenter retrospective international post-marketing cohort study including consecutive patients treated with SE for iAs between 2020 and 2022. Demographic, clinical, and angiographic data were collected. Primary effectiveness was independent core lab adjudicated complete occlusion rates (Raymond-Roy Class 1) at last follow-up. Primary safety were major ischemic/hemorrhagic events and mortality.
Results: In total, 305 patients with 332 aneurysms underwent SE implantation. The patients had a median age of 59 [50-67] years, and 256 (83.9%) were female. The baseline modified Rankin scale score was 0-2 in 291 patients (96.7%). Most aneurysms were unruptured (285, 93.4%) and saccular (309, 93.1%). Previous treatment was present in 76 (22.9%) patients. The median aneurysm size was 5.1 [3.4-9.0] mm, and the median neck width was 3.6 [2.7-5.1] mm. Most aneurysms were in the internal carotid artery C6 ophthalmic segment (126, 38.0%), followed by the communicating segment (58, 17.5%). At median 10.2 [6.4-12.9] months follow-up, 233 (73.0%) aneurysms achieved complete occlusion. After adjusting for confounders, complete occlusion remained consistent. Major stroke and procedure-related mortality were reported in 6 (2%) and 2 (0.7%) cases, respectively.
Conclusion: These results demonstrate that SE has a consistently high effectiveness and favorable safety for the treatment of iAs.
Source
Vivanco-Suarez J, Dibas M, Lopes DK, Hanel RA, Martínez-Galdámez M, Rodriguez-Calienes A, Cortez GM, Fifi JT, Devarajan A, Toth G, Patterson T, Altschul D, Pereira VM, Liu E, Puri AS, Kuhn AL, Guerrero WR, Khandelwal P, Bach I, Kan P, Edhayan G, Given C, Narayanan S, Gross BA, Farooqui M, Galecio-Castillo M, Derakhshani S, Ortega-Gutierrez S. Safety and effectiveness assessment of the surpass evolve (SEASE): a post-market international multicenter study. J Neurointerv Surg. 2024 Apr 25:jnis-2024-021503. doi: 10.1136/jnis-2024-021503. Epub ahead of print. PMID: 38663997.