We are upgrading the repository! A content freeze is in effect until December 6, 2024. New submissions or changes to existing items will not be allowed during this period. All content already published will remain publicly available for searching and downloading. Updates will be posted in the Website Upgrade 2024 FAQ in the sidebar Help menu. Reach out to escholarship@umassmed.edu with any questions.
A multiplex immunoassay using the Guthrie specimen to detect T-cell deficiencies including severe combined immunodeficiency disease
Document Type
Journal ArticlePublication Date
2010-09-28Keywords
Antigens, CD3Antigens, CD45
Blood Specimen Collection
Calibration
Feasibility Studies
Humans
Immunoassay
Immunologic Deficiency Syndromes
Infant, Low Birth Weight
Infant, Newborn
Leukocytes
Severe Combined Immunodeficiency
T-Lymphocytes
Genetics and Genomics
Medical Genetics
Pediatrics
Metadata
Show full item recordAbstract
BACKGROUND: Severe combined immunodeficiency (SCID) fulfills many of the requirements for addition to a newborn screening panel. Two newborn screening SCID pilot studies are now underway using the T-cell receptor excision circle (TREC) assay, a molecular technique. Here we describe an immunoassay with CD3 as a marker for T cells and CD45 as a marker for total leukocytes that can be used with the Guthrie specimen. METHODS: The multiplexing capabilities of the Luminex platform were used. Antibody pairs were used to capture and detect CD3 and CD45 from a single 3-mm punch of the Guthrie specimen. The assay for each biomarker was developed separately in identical buffers and then combined to create a multiplex assay. RESULTS: Using calibrators made from known amounts of leukocytes, a detection limit of 0.25 x 10(6) cells/mL for CD3 and 0.125 x 10(6) cells/mL for CD45 was obtained. Affinity tests showed no cross-reactivity between the antibodies to CD3 and CD45. The multiplex assay was validated against 8 coded specimens of known clinical status and linked to results from the TREC assay that had identified them. All were correctly identified by the CD345 assay. CONCLUSIONS: The performance parameters of the CD345 assay met the performance characteristics generally accepted for immunoassays. Our assay classifications of positive specimens concur with previous TREC results. This CD345 assay warrants evaluation as a viable alternative or complement to the TREC assay as a primary screening tool for detecting T-cell immunodeficiencies, including SCID, in Guthrie specimens.Source
Clin Chem. 2010 Sep;56(9):1460-5. Epub 2010 Jul 21. Link to article on publisher's siteDOI
10.1373/clinchem.2010.144329Permanent Link to this Item
http://hdl.handle.net/20.500.14038/43284PubMed ID
20660143Related Resources
Link to Article in PubMedae974a485f413a2113503eed53cd6c53