    A quantitative study of motion estimation methods on 4D cardiac gated SPECT reconstruction

    • CSV
    • RefMan
    • EndNote
    • BibTex
    • RefWorks
    Authors
    Qi, Wenyuan
    Yang, Yongyi
    Niu, Xiaofeng
    King, Michael A.
    UMass Chan Affiliations
    Department of Electrical and Computer Engineering
    Document Type
    Journal Article
    Publication Date
    2012-08-01
    Keywords
    Aged
    Algorithms
    Computer Simulation
    Female
    Heart
    Heart Ventricles
    Humans
    Image Processing, Computer-Assisted
    Male
    Middle Aged
    Models, Statistical
    Motion
    Myocardium
    Phantoms, Imaging
    Reproducibility of Results
    Time Factors
    Tomography, Emission-Computed, Single-Photon
    Radiology
    Show allShow less
    
    Metadata
    Show full item record
    Link to Full Text
    http://dx.doi.org/10.1118/1.4738377
    Abstract
    PURPOSE: Motion-compensated temporal processing can have a major impact on improving the image quality in gated cardiac single photon emission computed tomography (SPECT). In this work, we investigate the effect of different optical flow estimation methods for motion-compensated temporal processing in gated SPECT. In particular, we explore whether better motion estimation can substantially improve reconstructed image quality, and how the estimated motion would compare to the ideal case of known motion in terms of reconstruction. METHODS: We consider the following three methods for obtaining the image motion in 4D reconstruction: (1) the Horn-Schunck optical flow equation (OFE) method, (2) a recently developed periodic OFE method, and (3) known cardiac motion derived from the NURBS-based cardiac-torso (NCAT) phantom. The periodic OFE method is used to exploit the inherent periodic nature in cardiac gated images. In this method, the optical flow in a sequence is modeled by a Fourier harmonic representation, which is then estimated from the image data. We study the impact of temporal processing on 4D reconstructions when the image motion is obtained with the different methods above. For quantitative evaluation, we use simulated imaging with multiple noise realizations from the NCAT phantom, where different patient geometry and lesion sizes are also considered. To quantify the reconstruction results, we use the following measures of reconstruction accuracy and defect detection in the myocardium: (1) overall error level in the myocardium, (2) regional accuracy of the left ventricle (LV) wall, (3) accuracy of regional time activity curves of the LV, and (4) perfusion defect detectability with a channelized Hotelling observer (CHO). In addition, we also examine the effect of noise on the distortion in the reconstructed LV wall shape by detecting its contours. As a preliminary demonstration, these methods are also tested on two sets of clinical acquisitions. RESULTS: For the different quantitative measures considered, the periodic OFE further improved the reconstruction accuracy of the myocardium compared to OFE in 4D reconstruction; its improvement in reconstruction almost matched that of the known motion. Specifically, the overall mean-squared error in the myocardium was reduced by over 20% with periodic OFE; with noise level fixed at 10%, the regional bias on the LV was reduced from 20% (OFE) to 14% (periodic OFE), compared to 11% by the known motion. In addition, the CHO results show that there was also improvement in lesion detectability with the periodic OFE. The regional time activity curves obtained with the periodic OFE were also observed to be more consistent with the reference; in addition, the contours of the reconstructed LV wall with the periodic OFE were demonstrated to show less degree of variations among different noise realizations. Such improvements were also consistent with the results obtained from the clinical acquisitions. CONCLUSIONS: Use of improved optical flow estimation can further improve the accuracy of reconstructed images in 4D. The periodic OFE method not only can achieve improvements over the traditional OFE, but also can almost match that of the known motion in terms of the several quality measures considered.
    Source
    Med Phys. 2012 Aug;39(8):5182-93. doi: 10.1118/1.4738377. Link to article on publisher's site
    DOI
    10.1118/1.4738377
    Permanent Link to this Item
    http://hdl.handle.net/20.500.14038/48064
    PubMed ID
    22894443
    Related Resources
    Link to Article in PubMed
    ae974a485f413a2113503eed53cd6c53
    10.1118/1.4738377
    Scopus Count
    Collections
    Radiology Publications

    entitlement

    DSpace software (copyright © 2002 - 2023)  DuraSpace
    Lamar Soutter Library, UMass Chan Medical School | 55 Lake Avenue North | Worcester, MA 01655 USA
    Quick Guide | escholarship@umassmed.edu
    Open Repository is a service operated by 
    Atmire NV
     

    Export search results

    The export option will allow you to export the current search results of the entered query to a file. Different formats are available for download. To export the items, click on the button corresponding with the preferred download format.

    By default, clicking on the export buttons will result in a download of the allowed maximum amount of items.

    To select a subset of the search results, click "Selective Export" button and make a selection of the items you want to export. The amount of items that can be exported at once is similarly restricted as the full export.

    After making a selection, click one of the export format buttons. The amount of items that will be exported is indicated in the bubble next to export format.