An evaluation of iterative reconstruction strategies based on mediastinal lesion detection using hybrid Ga-67 SPECT images
Authors
Pereira, Nicholas F.Gifford, Howard C.
Pretorius, P. Hendrik
Smyczynski, Mark S.
Licho, Robert
Schneider, Peter B.
Farncombe, Troy H.
King, Michael A
UMass Chan Affiliations
Department of RadiologyDocument Type
Journal ArticlePublication Date
2008-11-01Keywords
AlgorithmsAnalysis of Variance
Gallium Radioisotopes
Humans
Image Processing, Computer-Assisted
Mediastinal Neoplasms
Tomography, Emission-Computed, Single-Photon
Radiology
Metadata
Show full item recordAbstract
Using psychophysical studies, the authors have evaluated the effectiveness of various strategies for compensating for physical degradations in SPECT imaging. The particular application was Ga-67-citrate imaging of mediastinal tumors, which was chosen because Ga-67 is a particularly challenging radionuclide for imaging. The test strategies included compensations for nonuniform attenuation, distance-dependent spatial resolution, and scatter applied in various combinations as part of iterative reconstructions with the rescaled block iterative-expectation maximization (RBI-EM) algorithm. The authors also evaluated filtered backprojection reconstructions. Strategies were compared on the basis of human-observer studies of lesion localization and detection accuracy using the localization receiver operating characteristics (LROC) paradigm. These studies involved hybrid images which were obtained by adding the projections of Monte Carlo-simulated lesions to disease-free clinical projection data. The background variability in these images can provide a more realistic assessment of the relative utility of reconstruction strategies than images from anthropomorphic digital phantoms. The clinical datasets were obtained using a GE-VG dual-detector SPECT system with CT-estimated attenuation maps. After determining a target lesion contrast, they conducted pilot LROC studies to obtain a near-optimal set of reconstruction parameters for each strategy, and then conducted the strategy comparison study. The results indicate improved detection accuracy with RBI-EM as more compensations are applied within the reconstruction. The relative rankings of the test strategies agreed in most cases with those of previous studies that employed simulated projections of digital anthropomorphic phantoms, thus confirming the findings of those studies.Source
Med Phys. 2008 Nov;35(11):4808-15. doi: 10.1118/1.2988164DOI
10.1118/1.2988164Permanent Link to this Item
http://hdl.handle.net/20.500.14038/48095PubMed ID
19070213Related Resources
Link to Article in PubMedae974a485f413a2113503eed53cd6c53