Modeling the performance characteristics of computed radiography (CR) systems
UMass Chan Affiliations
Department of RadiologyDocument Type
Journal ArticlePublication Date
2010-03-01Keywords
AlgorithmsComputer Simulation
Image Processing, Computer-Assisted
Linear Models
Mammography
Models, Theoretical
Monte Carlo Method
Radiographic Image Enhancement
Radiography
Bioimaging and Biomedical Optics
Biological and Chemical Physics
Biomedical Devices and Instrumentation
Diagnosis
Investigative Techniques
Medical Biophysics
Radiology
Metadata
Show full item recordAbstract
Computed radiography (CR) using storage phosphors is widely used in digital radiography and mammography. A cascaded linear systems approach wherein several parameter values were estimated using Monte Carlo methods was used to model the image formation process of a single-side read flying spot CR system using a granular phosphor. Objective image quality metrics such as modulation transfer function and detective quantum efficiency were determined using this model and show good agreement with published empirical data. A model such as that addressed in this work could allow for improved understanding of the effect of storage phosphor physical properties and CR reader parameters on objective image quality metrics for existing and evolving CR systems.Source