dc.identifier.citationHanel RA, Cortez GM, Lopes DK, Nelson PK, Siddiqui AH, Jabbour P, Mendes Pereira V, István IS, Zaidat OO, Bettegowda C, Colby GP, Mokin M, Schirmer CM, Hellinger FR, Given C, Krings T, Taussky P, Toth G, Fraser JF, Chen M, Priest R, Kan P, Fiorella D, Frei D, Aagaard-Kienitz B, Diaz O, Malek AM, Cawley CM, Puri AS, Kallmes DF. Prospective study on embolization of intracranial aneurysms with the pipeline device (PREMIER study): 3-year results with the application of a flow diverter specific occlusion classification. J Neurointerv Surg. 2023 Mar;15(3):248-254. doi: 10.1136/neurintsurg-2021-018501. Epub 2022 Mar 15. PMID: 35292570; PMCID: PMC9985759.
dc.identifier.issn1759-8478 (Linking)
dc.identifier.doi10.1136/neurintsurg-2021-018501
dc.identifier.pmid35292570
dc.identifier.urihttp://hdl.handle.net/20.500.14038/48624
dc.description<p>Full author list omitted for brevity. For the full list of authors, see article.</p>
dc.description.abstractBACKGROUND: The pipeline embolization device (PED; Medtronic) has presented as a safe and efficacious treatment for small- and medium-sized intracranial aneurysms. Independently adjudicated long-term results of the device in treating these lesions are still indeterminate. We present 3-year results, with additional application of a flow diverter specific occlusion scale. METHODS: PREMIER (prospective study on embolization of intracranial aneurysms with pipeline embolization device) is a prospective, single-arm trial. Inclusion criteria were patients with unruptured wide-necked intracranial aneurysms < /=12 mm. Primary effectiveness (complete aneurysm occlusion) and safety (major neurologic event) endpoints were independently monitored and adjudicated. RESULTS: As per the protocol, of 141 patients treated with a PED, 25 (17.7%) required angiographic follow-up after the first year due to incomplete aneurysm occlusion. According to the Core Radiology Laboratory review, three (12%) of these patients progressed to complete occlusion, with an overall rate of complete aneurysm occlusion at 3 years of 83.3% (115/138). Further angiographic evaluation using the modified Cekirge-Saatci classification demonstrated that complete occlusion, neck residual, or aneurysm size reduction occurred in 97.1%. The overall combined safety endpoint at 3 years was 2.8% (4/141), with only one non-debilitating major event occurring after the first year. There was one case of aneurysm recurrence but no cases of delayed rupture in this series. CONCLUSIONS: The PED device presents as a safe and effective modality in treating small- and medium-sized intracranial aneurysms. The application of a flow diverter specific occlusion classification attested the long-term durability with higher rate of successful aneurysm occlusion and no documented aneurysm rupture. TRIAL REGISTRATION: NCT02186561.
dc.language.isoen_US
dc.relation.urlhttps://doi.org/10.1136/neurintsurg-2021-018501
dc.rights© Author(s) (or their employer(s)) 2022. This is an open access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited, appropriate credit is given, any changes made indicated, and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/.
dc.rights.urihttp://creativecommons.org/licenses/by-nc/4.0/
dc.subjectaneurysm
dc.subjectflow diverter
dc.subjectintervention
dc.subjectCardiovascular Diseases
dc.subjectNeurology
dc.subjectNeurosurgery
dc.subjectRadiology
dc.titleProspective study on embolization of intracranial aneurysms with the pipeline device (PREMIER study): 3-year results with the application of a flow diverter specific occlusion classification
dc.typeJournal Article
dc.source.journaltitleJournal of neurointerventional surgery
dc.identifier.legacyfulltexthttps://escholarship.umassmed.edu/cgi/viewcontent.cgi?article=1704&amp;context=radiology_pubs&amp;unstamped=1
dc.identifier.legacycoverpagehttps://escholarship.umassmed.edu/radiology_pubs/687
dc.identifier.contextkey28980191
refterms.dateFOA2022-08-23T17:21:53Z
html.description.abstract<p>BACKGROUND: The pipeline embolization device (PED; Medtronic) has presented as a safe and efficacious treatment for small- and medium-sized intracranial aneurysms. Independently adjudicated long-term results of the device in treating these lesions are still indeterminate. We present 3-year results, with additional application of a flow diverter specific occlusion scale.</p> <p>METHODS: PREMIER (prospective study on embolization of intracranial aneurysms with pipeline embolization device) is a prospective, single-arm trial. Inclusion criteria were patients with unruptured wide-necked intracranial aneurysms < /=12 mm. Primary effectiveness (complete aneurysm occlusion) and safety (major neurologic event) endpoints were independently monitored and adjudicated.</p> <p>RESULTS: As per the protocol, of 141 patients treated with a PED, 25 (17.7%) required angiographic follow-up after the first year due to incomplete aneurysm occlusion. According to the Core Radiology Laboratory review, three (12%) of these patients progressed to complete occlusion, with an overall rate of complete aneurysm occlusion at 3 years of 83.3% (115/138). Further angiographic evaluation using the modified Cekirge-Saatci classification demonstrated that complete occlusion, neck residual, or aneurysm size reduction occurred in 97.1%. The overall combined safety endpoint at 3 years was 2.8% (4/141), with only one non-debilitating major event occurring after the first year. There was one case of aneurysm recurrence but no cases of delayed rupture in this series.</p> <p>CONCLUSIONS: The PED device presents as a safe and effective modality in treating small- and medium-sized intracranial aneurysms. The application of a flow diverter specific occlusion classification attested the long-term durability with higher rate of successful aneurysm occlusion and no documented aneurysm rupture.</p> <p>TRIAL REGISTRATION: NCT02186561.</p>
dc.identifier.submissionpathradiology_pubs/687
dc.contributor.departmentRadiology


Files in this item

Thumbnail
Name:
248.full.pdf
Size:
1.234Mb
Format:
PDF
Description:
Most recent version 2023
Thumbnail
Name:
neurintsurg_2021_018501.full.pdf
Size:
1.234Mb
Format:
PDF
Description:
Previous version (15 March 2022)

This item appears in the following Collection(s)

Show simple item record

© Author(s) (or their employer(s)) 2022. This is an open access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited, appropriate credit is given, any changes made indicated, and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/.
Except where otherwise noted, this item's license is described as © Author(s) (or their employer(s)) 2022. This is an open access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited, appropriate credit is given, any changes made indicated, and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/.